Setting up | Connecting | Measuring | Timebase | Y amplifier | AC/GND/DC
Next Page: Power Supplies
Also See: AC, DC and Electrical Signals
An oscilloscope is a test instrument which allows you to look at the 'shape' of electrical signals by displaying a graph of voltage against time on its screen. It is like a voltmeter with the valuable extra function of showing how the voltage varies with time. A graticule with a 1cm grid enables you to take measurements of voltage and time from the screen.
The graph, usually called the trace, is drawn by a beam of electrons striking the phosphor coating of the screen making it emit light, usually green or blue. This is similar to the way a television picture is produced.
Oscilloscopes contain a vacuum tube with a cathode (negative electrode) at one end to emit electrons and an anode (positive electrode) to accelerate them so they move rapidly down the tube to the screen. This arrangement is called an electron gun. The tube also contains electrodes to deflect the electron beam up/down and left/right.
The electrons are called cathode rays because they are emitted by the cathode and this gives the oscilloscope its full name of cathode ray oscilloscope or CRO.
A dual trace oscilloscope can display two traces on the screen, allowing you to easily compare the input and output of an amplifier for example. It is well worth paying the modest extra cost to have this facility.
Oscilloscope circuit symbol
Oscilloscope
Photograph © Rapid Electronics
Rapid Electronics: Oscilloscopes
Oscilloscopes are complex instruments with many controls and they require some care to set up and use successfully. It is quite easy to 'lose' the trace off the screen if controls are set wrongly!
The picture shows what you should see after setting up, when there is no input signal connected. There is some variation in the arrangement and labelling of the many controls so the instuctions may need to be adapted for your instrument.
Further information on the controls: Timebase | Y amplifier | AC/GND/DC
The Y INPUT lead to an oscilloscope should be a co-axial lead and the diagram shows its construction. The central wire carries the signal and the screen is connected to earth (0V) to shield the signal from electrical interference (usually called noise).
Construction of a co-axial lead
Most oscilloscopes have a BNC socket for the y input and the lead is connected with a push and twist action, to disconnect you need to twist and pull. Oscilloscopes used in schools may have red and black 4mm sockets so that ordinary, unscreened, 4mm plug leads can be used if necessary.
Professionals use a specially designed lead and probes kit for best results with high frequency signals and when testing high resistance circuits, but this is not essential for simpler work at audio frequencies (up to 20kHz).
An oscilloscope is connected like a voltmeter but you must be aware that the screen (black) connection of the input lead is connected to mains earth at the oscilloscope. This means it must be connected to earth or 0V on the circuit being tested.
Oscilloscope lead and probes kit
Photograph © Rapid Electronics
If you are using an oscilloscope for the first time it is best to start with an easy signal such as the output from an AC power pack set to about 4V. The picture shows the trace you should see after setting the controls correctly.
After connecting the oscilloscope to the circuit you wish to test you will need to set the controls to obtain a clear and stable trace on the screen:
Further information on the controls: Timebase | Y amplifier | AC/GND/DC
The trace on an oscilloscope screen is a graph of voltage against time. The shape of this graph is determined by the nature of the input signal.
In addition to the properties labelled on the graph, there is frequency which is the number of cycles per second.
The diagram shows a sine wave but the properties apply to any signal with a constantly repeating shape.
Frequency and time period are the inverse of each other:
frequency = | 1 |
time period |
and
time period = | 1 |
frequency |
The picture shows the trace of an AC signal on an oscilloscope with the controls on these settings:
Some low cost oscilloscopes have small screens where the grid lines are less than 1cm apart. On these instruments the controls will be marked '/div' (per division) instead of '/cm'.
Voltage is shown on the vertical y-axis and the scale is determined by the Y AMPLIFIER (VOLTS/CM) control. Usually peak-peak voltage is measured because it can be read correctly even if the position of 0V is not known. The amplitude is half the peak-peak voltage.
Voltage = distance in cm × volts/cm |
For the example trace shown above:
peak-peak voltage = 4.2cm × 2V/cm = 8.4V
amplitude (peak voltage) = ½ × peak-peak voltage = 4.2V
If you wish to read the amplitude directly from the screen you must first check the position of 0V (normally halfway up the screen). Move the AC/GND/DC switch to GND (0V) and use Y-SHIFT (up/down) to adjust the position of the trace if necessary. Switch back to DC afterwards so you can see the signal again.
Time is shown on the horizontal x-axis and the scale is determined by the TIMEBASE (TIME/CM) control. The time period (often just called period) is the time for one cycle of the signal. The frequency is the number of cyles per second, frequency = 1/time period.
Ensure that the variable timebase control is set to 1 or CAL (calibrated) before attempting to take a time reading.
Time = distance in cm × time/cm |
For the example trace shown above:
time period = 4.0cm × 5ms/cm = 20ms
frequency = 1/time period =
1/20ms = 50Hz
The oscilloscope sweeps the electron beam across the screen from left to right at a steady speed set by the TIMEBASE control. Each setting is labelled with the time the dot takes to move 1cm, effectively it is setting the scale on the x-axis. The timebase control may be labelled TIME/CM.
At slow timebase settings (such as 50ms/cm) you can see a dot moving across the screen, as in the upper picture.
At fast timebase settings (such as 1ms/cm) the dot is moving so fast that it appears to be a line, as in the lower picture.
The VARIABLE timebase control can be turned to make a fine adjustment to the speed, but it must be left at the position labelled 1 or CAL (calibrated) if you wish to take time readings from the trace drawn on the screen.
The TRIGGER controls are used to maintain a steady trace on the screen. If they are set wrongly you may see a trace drifting sideways, a confusing 'scribble' on the screen, or no trace at all. The trigger maintains a steady trace by starting the dot sweeping across the screen when the input signal reaches the same point in its cycle each time.
For straightforward use it is best to leave the trigger level set to AUTO, but if you have difficulty obtaining a steady trace try adjusting this control to set the level manually.
The oscilloscope moves the trace up and down in proportion to the voltage at the Y INPUT and the setting of the Y AMPLIFIER control. This control sets the voltage represented by each centimetre (cm) on the the screen, effectively it is setting the scale on the y-axis. Positive voltages make the trace move up, negative voltages make it move down.
The picture shows a varying DC signal which is always positive.
The y amplifier control may be labelled Y-GAIN or VOLTS/CM.
The input voltage moving the dot up and down at the same time as the dot is swept across the screen means that the trace on the screen is a graph of voltage (y-axis) against time (x-axis) for the input signal.
The normal setting for this switch is DC for all signals - including AC signals!
Switching to GND (ground) connects the y input to 0V and allows you to quickly check the position of 0V on the screen. This is normally halfway up, as shown in the picture. There is no need to disconnect the input lead while you move the switch to GND because the input is disconnected internally.
Switching to AC inserts a capacitor in series with the input to block out any DC signal present and pass only AC signals. This is used to examine signals showing a small variation around one constant value, such as the ripple on the output of a smooth DC supply. Reducing the VOLTS/CM to see more detail of the ripple would normally take the trace off the screen! The AC setting removes the constant (DC) part of the signal, allowing you to view just the varying (AC) part which can now be examined more closely by reducing the VOLTS/CM. This is shown in the diagrams below:
1. Switch in normal DC position.
The ripple is difficult to see clearly but if VOLTS/CM is reduced to try enlarge the ripple, the trace will disappear off the screen.
2. Switch moved to AC position.
The constant (DC) part of the signal has been removed, leaving just the ripple (AC) part.
3. VOLTS/CM reduced to enlarge the ripple.
The ripple can now be examined more closely.
Next Page: Power Supplies | Study